Microtips
Microtip probes are
available in tapered and stepped
configurations and are designed
to process samples contained in
small diameter vessels at extremely
high intensity (refer to chart for
recommended volumes).
Tapered microtips screw into
standard 1/2" step horns with
tapped ends.
The double stepped microtip
consists of a coupler and a stepped
tip and screws directly into the
converter. It has lower amplitude
than the tapered tip, but is capable
of reaching into small diameter
vessels of greater depth and can
process solutions down to 0.1 ml.
The horns and microtips detailed
on this sheet are also compatible
with the earlier Branson Sonifier,
Model 350. For information
on horns for Sonifier Models
185-200, call us or your Customer
Service Representative.
WARNING: Because of their
high amplitude, microtips are
highly stressed and can break
if operated in air or at an output
control setting greater than 7
on all Sonifier models.